• Home
  • Contact
  • Sitemap
  • Imprint & Privacy Policy
PSS

PSS - Polmer Standards Service

Expertise and Reliability for every step of Polymer Characterization

  • Company
  • Products
  • Services
  • Solutions
  • InfoCenter
  • NetCommunity
 
  • Polymer Standards and Reference Materials
  • Columns for GPC,SEC,GFC
  • (Macromolecular) Chromatography Data Systems
  • LC Instruments and Detectors
  • Contract Analysis (for Polymer Analysis & Testing)
  • Site Search for Products, Applications and Publications
 

Specifications GRAM columns

Applicability

Polyurethane, Polyimide, Starches, Celluloses, certain Polyamide, other polar polymers

Eluents

DMF, DMAc, NMP, DMSO

Material

Polyester copolymer network

Maximum Pressure

50 -120 bar (725 - 1740 psi), depending on porosity

Maximum Temperature

90° C

Maximum Flow Rate

2 ml/min (I.D.: 8 mm)

Particle Size

10 µm

Molecular Weight Range

100 to 50 000 000 Da

Incompatibility

Drying and freezing

Recommended Column Combination

Low molar mass:     100 Å +     100 Å 

Medium molar mass:  30 Å +   1 000 Å +   1 000 Å

High molar mass:      100 Å + 10 000 Å + 10 000 Å

 

 

Separation range

30 Å

100

 - 

10 000 Da

100 Å

300

 - 

60 000 Da

1 000 Å

1 000

 - 

1 000 000 Da

3 000 Å

 5 000

 - 

5 000 000 Da

10 000 Å 

 10 000

 - 

50 000 000 Da

linear

500

 - 

1 000 000 Da

 

 

Part numbers

Calibration curves: single porosity and linear columns

Calibrants: Poly(methyl methacrylate)" rel="fancybox110" class="jqfancybox">
Eluent: DMF
Calibrants: Poly(methyl methacrylate)

 

 

< Back to GPC/SEC columns - Powerful Solutions for Size Separation

 

Search


  • New brochure GPC/SEC Services and Contract Analysis

    Comprehensive Macromolecular Characterization

    more

  • Academic Promotion Program EcoSEC

    starting June 01 2010

    more

  • GreenGPC/SEC - Tour 2010 Europe

    PSS and Tosoh Bioscience invite you to a free-of-charge 1-day seminar:

     

    - March...

    more